Dialysis Machine MSLHM04 Single Pumps LCD Touch Screen Hemodialysis Machine

Apply to hemodialysis, hem perfusion, ISO UF, plasma exchange, etc.

Technical Feature

  • Body Fluids Processing Equipments
  • MSL
  • MSLHM04
  • Guangdong, China (Mainland)
  • 10" LCD touch screen
  • 5℃~30℃
Send Project Inquiry Add Inquiry List

0 Add Favorite

DESIGN SERVICE OVERVIEW

Project Design Scope

Project Service Scope

PRIMARY PARAMETERS

DETAIL DESCRIPTION

Product Description

Technical Specification of MSLHM04 Single Pumps LCD Touch Screen Hemodialysis Machine

Apply to hemodialysis, hem perfusion, ISO UF, plasma exchange, etc.
User friendly operation
All-around security system
Precise UF system
Powerful functions
User friendly operation
10” LCD touch screen
Automatically activates treatment program when blood is detected flowing through
Auto self-test, prime, rinse, disinfection, hot disinfection, treatment and power off, etc.
Records in table and graphs
All-around security system
Conductivity
Temperature
Air bubble & Blood level
Blood leakage
Blood pressure
Arterial, Venous & TM Pressure
Anti-coagulation
Backup battery
Precise UF system
Classic waterway system controls the UF precisely
Double circulation waterway system to prevent cross-infection
Powerful functions
Na+, UF & HCO3- Profiles
Suitable for any dialysate formula
Dialysate flow linearly adjustable
Kt/V
Optional: Bicarb-cartridge holder, Pyrogen filter port, Network connection, etc.

◆ Volume (length× width ×height): 370mm×340mm ×1420mm
◆ Weight: about 90kg   
◆ Power supply voltage: AC220V±10%   
·Frequency: 50HZ ~ 60 HZ   ·Power: 1700W
◆ Blood pump:
·Flux: 10 ~ 460 ml/min (Ф 6.4mm)
·Flux: 20 ~ 620 ml/min (Ф 8mm)
◆ Heparin pump:
·Flux: 0.1ml/h~10ml/h(±5%)
·Precise: 0.1ml/h
·Injector size: 10ml/20ml/30ml(selectable)
◆ Arterial pressure:

display scope:-300mmHg~+400mmHg(±10mmHg)
◆ Venous pressure:
display scope:-200mmHg~+500mmHg(±10mmHg)
◆ Dialysate flow:
·300ml/min~800ml/min linearity adjustable(+10%)
·Temperature:35.0℃~39.0℃
·Resolving rate:0.1℃
·Conductivity:12.3mS/cm~15.8mS/cm(±0.1 mS/cm)
◆ UF flow scope: 0~4000ml/h
◆ ISO UF flow scope:0~4000ml/h(±30ml/h)
◆ TMP Scope:-100mmHg~+600mmHg(±20mmHg)
◆ Blood leakage monitor:

over 1ml blood per liter dialysate(flow:500ml/min)
◆ Blood level monitor: ultrasonic sensor
◆ Air bubble monitor: Infrared and response threshold value:
Single air bubble of 200uL exists when blood flux is 200ml/min
◆ Inflow pressure: 0.12MPa~0.6MPa, 0MPa~0.6MPa (Optional)
◆ Inflow temperature: 5℃~30℃
◆ Environment temperature: 10℃~38℃, Relative humidity≤80%
◆ Rinse/disinfection: chemical disinfection (citric acid, peracetic acid)
◆ Hot rinse: ≥80℃
◆ Back-up power supply: last for 15-30min after electric-cut.

Detailed Images
Packing & Delivery
Our Company

Send Consultations Sending PUBLIC question and getting response from the supplier

Send Only your public informations will be shown

All Consultations