LDY1005 Laser Particle Size Analyzer works with principle of Mie Scattering to precisely determine the particle size distribution from 0.1 to 500 micrometer. This equipment has a few advantages all its own. Robust laser diode and innovative photocells optimize the precision of static light scattering and give LDY1005 a longer operational lifetime. Patented optical bench uproots light path deviation, one of the major technical problems in the industry of particle size analysis. User-friendly program and advanced algorism bring accuracy and repeatability qualification to International Standard ISO13320-1 Particle Size Analysis - Laser Diffraction Methods. LDY1005 has become the most reliable and cost-effective LPSA and being widely used in Particle Characterization.
Specifications
Measurement Principle: Mie Scattering Theory/Laser Diffraction Principle
Size Range: 0.1-500 um ( particle size D50, D03, D06, D10, D16, D25, D75, D84, D90, D97, D98 etc. )
Repeatability: <1% (Certified Reference Material SB2005 D50)
Accuracy: <1% (Certified Reference Material SB2005 D50)
Light Source: Semiconductor Laser Diode(life time more than 25000 hrs)
Measurement Time: <60 seconds (Sample preparation time is not included)
Power: 220V 50Hz 80W(Power converter 110V-220V is availabe as per specific requirement)
Data Connection: RS-232
Operating Systems: Windows XP/ Windows 7
ISO Norms: ISO 13320
Demo Videos
Particle Size Analyzer (Small Cell)
Particle Size Analyser (Dispersing Unit)
Applications
Particle size distribution such as Median D50, D03, D06, D10, D16, D25, D75, D84, D90, D97,D98 of Abrasives, Adhesives, Agrochemical, Barite, Batteries, Bentonite, Boron Carbide, Brucite, Bubble, Calcite, Calcium Carbonate, Carbon Black, Catalysts, Cement, Ceramics, Chemicals, Clay, Coal, Coatings, Corundum, Cosmetics, Diamond Powder, Dolomite, Diatomite, Emulsion, Environmental, Explosives, Ferrite, Flour, Fluorescent, Fluorite, Food & Beverage, Food Additive, Graphite, Grinding, Inks, Kaolin, Medicine, Metal Powder, Mica, Milling, Minerals, Oxides, Paints, Paper, Petrochemical, Pharmaceuticals, Pigments, Plaster, Plastics, Polymers, Quartz, Refractory, Resins, Silica, Slurries, Soil Sediments, Starch, Sulfur, Synthetics, Talc, Toners, Tourmaline, Wollastonite, Zeolite, Zirconium Silicate etc in micron range. The particle size analysis of above materials are easy for LDY1005 laser particle size analyzer who meets specific requirement of both research and quality control with its qualified performance conforming to ISO 13320 Particle Size Analysis-Laser Diffraction Methods.
Special Coating
External (corrosion resistant) and internal (anti-reflective) coatings effectively shield the interference by unwanted light and reflecting walls. Special coating protects the laser diffraction and improves the analyzer performance.
Robust Light Source
Excellent monochromatic laser diode works for at least 25,000 hours. No work, no loss. Monochrome improves the accuracy and long lifetime extends the working life of laser particle size analyzer. A novel technology integrates Fourier lens into the holder of laser diode which optimizes the light path alignment.
Stable Optical Bench
All major optical assemblies such as laser, cell holder and detection system are firmly fixed on one cast steel optical bench. This patented design uproots light path deviation, one of the major technical problems in the industry of particle size analysis.
Sensitive Photocells
High-resolution photocells make LDY1005 much more sensitive than any other particle size analyzer. Detection system array effectively captures the scattered light thus ensures the authenticity and accuracy of particle size characterization.
High-Speed Circuit Boards
Circuit boards of serial port, filtering, amplification and data transmit & process have a considerable advantage in speed over traditional ones. They work with new algorism to give a satisfactory measurement time.
Versatile Small (Volume) Cell Assembly
The small cell assembly has found wide applications because of its very small fault rate and high reliability. Corrosion-resistant small cell can either hold water or other mediums. Micro silent agitator fully stirs up the suspension and improves the accuracy and repeatability.
Optional Dispersing Unit
Dispersing Unit integrates functions of dispersing, agitating and circulating. It effectively suspends high-density particles in water to achieve a better accuracy. The panel of HYL-300 is simple and easy to use.
Ingenious Cell Holder
Patented cell holder makes the shift between small cell and flow cell assembly easy and safe.
Advanced Mie Scattering
LDY1005 laser particle size analysis system adopts the most advanced principle of Mie Scattering and gives the micron particles a scientific and objective particle size analysis. Theory of particle size analyzer is very important
Accurate & Fast Algorism
A new algorism has been designed which will be able to obtain the accurate results and greatly reduce the measurement time.
Flexible Parameters
In order to improve the correlation capability of LDY1005 with other laser particle size analyzers, Labmen instrument limited develops advanced and flexible parameters to make LDY1005 the preferred substitute of world renowned. These parameters can provide users with the acceptable correlation results.
Safe Log on
User can log on particle sizing software in two ways: Administrator or Operator, only Administrator is authorized to modify the major settings of particle size analysis system, which greatly improves the security and stability.
Automatic Measurement
The automation starts with automatic error detecting on background and light path, and then size analysis is carried out automatically. In less than 60 seconds size distribution is displayed on the computer screen ready for exporting to Excel or print.
Comprehensive Analysis report
The result analysis report includes typical values, size distribution curve/histogram as well as table of size distribution in order to compare with other results respectively. Particle size includes D50, D03, D06, D10, D16, D25, D75, D90, D97, D98. User is also allowed to edit and customize the report.
Authoritative Validation & Traceability
Labmen instrument Limited validate each unit by measuring stable Certified Reference Material CRM SB2005, which ensures the performance of each analyzer and optimizes the traceability. The results obtained must agree with the standard values before export.
Satisfying Accuracy & Repeatability
More than meeting the requirement of ISO13320-1 on accuracy and repeatability (<3%), LDY1005 further optimizes the reproducibility to a new level (<3%).
LDY1005 Laser Particle Size Analyzer | |
Measurement Principle | Mie Scattering Theory/Laser Diffraction Principle |
Size Range | 0.1-500 um ( particle size D50, D03, D06, D10, D16, D25, D75, D84, D90, D97, D98 etc. ) |
Repeatability | <1% |
Accuracy | <1% |
Light Source | Semiconductor Laser Diode(life time more than 25000 hrs) |
Measurement Time | <60 seconds |
Power | 220V 50Hz 80W |
Data Connection | RS-232 |